- 48 -
UNIT
DIRECTION
CALIBRATION
UNIT
DIRECTION
CALIBRATION
(5) Impact Direction Setting
1>Press key to enter the menu;
2>Press key to select DIRECTION and then
press key to enter the setting page. Press
key to set the impact direction and then press
key to confirm.
3>Press key to return to the main interface.
(6) Self-calibration
1>Press key to enter the menu;
2>Press key to select CALIBRATION, and then
press key to enter the CALIBRATION page.
Select different test points, after 5 times consecutive
load impact, according to the measured average
value and measured value, referring to the standard
hardness value, adjust the 3 digit values of the
measured value through key, and then press
key to complete calibration.
- 49 -
HARDNESS / STRENGTH
LANGUAGE
HARDNESS / STRENGTH
LANGUAGE
3>Press key to return to the main interface.
(7) HARDNESS/STRENGTH Setting
1>Press key to enter the menu;
2>Press key to select HARDNESS/STRENGTH
and then press key to enter the setting page. Press
key to choose HARDNESS or STRENGTH and
then press key to confirm.
3>Press key to return to the main interface.
(8) LANGUAGE Setting
1>Press key to enter the menu;
2>Press key to select LANGUAGE and then
press key to enter the setting page. Press
key to choose 中文 or ENGLISH and then press
key to confirm.
3>Press key to return to the main interface.
- 50 -
● The function of auto-off for saving power; the device
will turn itself off without any operation in five minutes;
● On screen shows LOW signal when power is low,
and shut it down automatically.
7.4 Auto-off
9.Maintenance
● Clean the conduit and the impact body of the impact
device with nylon brush after utilization 1000~2000
times. First screw the back-up ring off, pick the impact
body out, screw nylon brush inside conduit
counterclockwise till the end, then pull it out. Repeat 5
times, then put the impact body and back-up ring on;
● Release the impact body after using;
● All kinds of lubricant are not allowed inside the impact
device.
9.1 About the Impact Device
8.Malfunction Analysis and
Troubleshooting
Problem
description Cause Solution
Change batteries The battery
is exhausted
Power-on
is failed
- 51 -
● Testing the standard Rockwell hardness block, all errors
are more than 2HRC, the possible reason is the
wear-out-failure of the head, changing it or the impact
body should be considered;
● When the tester appears other abnormal phenomenon,
please do not remove or adjust any fixed assembly
parts. Please fill out the warranty card and submit it to
our maintenance department to implement the warranty
regulations.
● The product is powered by 3.7V 350mAh lithium
battery, which is not removable. When the product
cannot be turned on or the power display is blank,
please charge it in time. Please use the DC 5V 0.5A
charging adapter to charge the product, and the
charging interface is Type-C.
● When charging, battery symbol scrolls through. When
fully charged, the battery symbol is displayed as full
battery .
● Store it in normal temperature and far away vibration,
strong magnetic field,corrosive medium, humidity and
dust;
● Transport it with original package in three-class
highway.
9.2 About the Device
9.3 About the Battery
Attention:
When not used for a long time, fully charge the product
and recharge it every six months to avoid battery damage.
9.4 Storage Condition, Transportation
and Tips
- 52 -
Material C G E DL
Steel/
Cast steel
HRC
HRC
HRB
HRB
HRB
HRB
HB
HV
HV
HV
HS
HB
HB
HB
HB
HB
HB
HB
HB
HRA
Forged
steel
Gray Cast
iron
Nodular Cast
iron
Cast
aluminum
alloy
Copperzinc alloy
(Brass)
Copper-tin
alloy (Bronze)
Copper
17.9~68.5 19.3~67.9
59.6~99.6
59.1~85.8
127~651
83~976
32.2~99.5
143~650
20.4~67.1
80~898
46.5~101.7
85~655
85~802
93~334
131~387
19~164
23.8~84.6
40~173
13.5~95.3
60~290
45~315
23~210
22.7~85.0 23.8~85.5
127~364
32~168
92~326
80~935 100~941 82~1009
19.8~68.2 20.7~68.2
80~638
80~937
33.3~99.3
80~683
80~996
31.8~102.1
90~646 83~663
84~1042
35.8~102.6
22.6~70.2
81~646
61.7~88.0
47.7~99.9 37.0~99.9
80~950
30.6~96.8
20.0~69.5 22.4~70.7 20.6~68.2
Impact device Unit
Alloyed
Tool steel
Stainless
steel
D/DC D+15
Attached Table 1
- 53 -
No. HLD σ b(MPa)
1
2
3
4
5
6
7
8
9
10
350~522 374~780
737~1670
707~1829
704~1980
763~2007
721~1875
755~1993
703~1676
1180~2652
844~1933
500~710
500~730
500~750
500~750
500~738
540~738
500~750
630~800
500~710
mild steel
high carbon steel
chrome steel
Cr-V steel
Cr-Ni steel
Cr-Mo steel
Cr-Ni-Mo steel
Cr-Mn-Si steel
SSST steel
SST steel
Material
Attached Table 2
- 54 -
Attached Table 3
the range of impact device application:
DC-type for hole or inner cylindrical surface;
DL-type for spindly narrow groove or hole;
D-type for regular testing;
D+15 for groove or sunken surface due to its smaller contact area and lengthening;
C-type for hardened layer and samples with minor diameter and thinner wall due to
its smaller impact force which usually cause damage on surface;
G-type for larger thickness and more rough casting samples;
E-type for highest hardness materials.
Shaped impact
device DC(D)/DL D+15 C G E
impact energy 11mJ 11mJ 2.7mJ 90mJ 11mJ
impact body quality 5.5g/7.2g 7.8g 3.0g 20.0g 5.5g
bulb-head hardness 1600HV 1600HV 1600HV 1600HV 5000HV
bulb-head diameter 3mm 3mm 3mm 5mm 3mm
bulb-head material tungsten
carbide
tungsten
carbide
tungsten
carbide
tungsten
carbide adamas
impact device
diameter 20mm 20mm 20mm 30mm 20mm
impact device
length
86(147)/
75mm 162mm 141mm 254mm 155mm
impact device
quality 50g 80g 75g 250g 80g
sample max
hardness 940HV 940HV 1000HV 650HV 1200HV
bulb-head indentation dimension
sample surface
mean roughness 1.6µm 1.6µm 0.4µm 6.3µm 1.6µm
sample min quality
testing directly
stable support
tight coupling
>5kg
2~5kg
0.05~2kg
>5kg
2~5kg
0.05~2kg
>1.5kg
0.5~1.5kg
0.02~0.5kg
>15kg
5~15kg
0.5~5kg
>5kg
2~5kg
0.05~2kg
sample min thickness
tight coupling
harden layer min depth
5mm
≥0.8mm
5mm
≥0.8mm
1mm
≥0.2mm
10mm
≥1.2mm
5mm
≥0.8mm
hardness 300HV
indentation diameter
indentation depth
0.54mm
24µm
0.54mm
24µm
0.38mm
12µm
1.03mm
53µm
0.54mm
24µm
hardness 600HV
indentation diameter
indentation depth
0.54mm
17µm
0.54mm
17µm
0.32mm
8µm
0.90mm
41µm
0.54mm
17µm
hardness 800HV
indentation diameter
indentation depth
0.35mm
10µm
0.35mm
10µm
0.35mm
7µm
---
---
0.35mm
10µm
- 55 -
Attached Table 4
Model No. Remark Shaped back-up
ring sketch No.
1 Z10-15
Z14.5-30
Z25-50
HZ11-13
HZ12.5-17
HZ16.5-30
K10-15
K14.5-30
HK11-13
HK12.5-17
HK16.5-30
UN
test outside cylindrical
surfaceR10~R15
test outside cylindrical
surfaceR14.5~R30
test outside cylindrical
surfaceR25~R50
test inner cylindrical
surfaceR11~R13
test inner cylindrical
surfaceR12.5~R17
test inner cylindrical
surfaceR16.5~R30
test outside sphere
surfaceSR10~SR15
test outside sphere
surfaceSR14.5~SR30
test inner sphere
surfaceSR11~SR13
test inner sphere
surfaceSR12.5~SR17
test inner sphere
surfaceSR16.5~SR30
test outside cylindrical
surface, adjustable
radius R10~∞
2
3
4
5
6
7
8
9
10
11
12
Sndway Technology (Guangdong) Co., LTD
全国咨询服务热线/ Service Hotline:400-125-6969
深达威科技(广东)股份有限公司
Add: Sndway Science & Technology Industrial Park, 58
Tuanjie Road, Humen 523930, Dongguan, China
地 址:东莞市虎门镇虎门团结路58号深达威科技园
电 话 / Tel:0769-85265688
网 址/ Web:www.sndway.com
邮 箱/E-mail:market@sndway.com