- 46 -
6. Detailed Operation Guide
6.2 Instrument Description
6.1 Power on/off
Power on: Long press to start the device;
Power off: Long press to turn the device off;
Auto-off: The device will automatically shut down without
operations within 5mins.
Forced shutdown: Long press for more than 10s to shut
down it.
The device will automatically enter the main display interface
after starting, as shown in the figure below:
08:28
DL HL
Avg: 789HL
Mat: Brass
Cnt: 03
Current test
value 789
Type of probe
Impact direction
Average value
Impact times
Time Display
Hardness unit
Material
Battery level
Light indicator
Alarm ring
- 47 -
● Long press to store the current data, which can only be
saved once. It is valid only after the set impact times is reached.
● Press to delete the last single measurement value, but
confirm it in the following display interface:
(1) Description of Main Display Interface
(2) Operation of Measurement
(3) Operation of Keys
Time Display: Display the current time.
Battery level: Display the remaining power.
Impact direction: Current impact direction.
Average value: Current avarage value.
Hardness unit: Hardness unit of current test value.
Test value: Current single test value.
Material: Current set material.
Impact times: Displays the impact times that have been
completed. When viewing a single measurement value, the
screen displays the corresponding times of the single
measurement.
Measurement can be carried out under the main interface. After
each measurement is completed, the measured value will be
displayed, and the number of impact will be increased by 1. If
the tolerance limit is exceeded, the instrument will \"beep\" once,
after reaching the set impact times, the instrument will \"beep\"
twice.
- 48 -
Are you sure you
want to delete?
Yes No
● Press to change the material, which can be circularly
selected from a variety of alternative materials. If the current
setting is hardness measurement, the hardness unit will be
changed to HL every time user press . So user should set
the material first and then the hardness unit when measuring.
1> Press to move the cursor to [Yes] and then short
press to confirm deletion of the last single measurement
value.
2> Press to move the cursor to [No] and then short
press to cancel the deletion.
● In the main screen, long press to enter the main menu;
● Short press to change the impact direction;
● Short press can change the hardness unit, it can
converse in a variety of hardness units circularly;
3> Regardless of the cursor position, pressing can cancel
the deletion, too.
Note:
The so-called \"transformation\" refers to the corresponding
relationship established for a material according to the Richter
hardness and other hardness on the basis of a large number of
tests. Based on this relationship, the hardness tester
automatically converts the measured Leeb hardness value to
the hardness value of other hardness unit.
- 49 -
6.3 Menu Structure Chart
The setting of instrument parameters and additional functions
can be realized through menu operation.
Main screen
MEAS Setting Record SYS Setting Calibration Information RES default
1> Press to move cursor upward;
2> Press to move cursor downward;
3> Press to move cursor leftward;
4> Press to move cursor rightward;
5> Press to enter options;
6> Press to back to the main screen.
Long press on the main screen to
enter the menu. The instrument has six
menu options: MEAS Setting, Record,
SYS Setting, Calibration, Information,
RES default.
The specific operations are as follows:
6.4 Menu Operations
MEAS Setting Record
SYS Setting Calibration
Information RES default
08:28
The first item
Delete all
Delete item
The current item
The last item
Impact DIR
Hard/strength
Tolerance
Hardness Unit
Material
Average Times
Auto Save:
Remove error:
Key Tone:
Alarm Tone:
Light Set:
ON
OFF
ON
ON
ON
Brightness
Language
Date Time Set
- 50 -
Enter the menu, select measurement condition , and press
to enter the options, refer to <6.4 Menu Operation>;
The measurement conditions include: Impact DIR, Average Times,
Material, Hardness Unit, Tolerance limit, Hardness/strength.
1> Press to move cursor upward;
2> Press to move cursor downward;
3> Press to enter the option;
4> Press to back to the main menu.
Operations are as follows:
6.5 Testing Condition Setting
(1) Setting of Impact Direction
Note:
When [strength] pattern is set, user can not set [Hardness Unit]
pattern, so the cursor will automatically skip the [Hardness Unit]
pattern.
1> Press to move the cursor
to the impact direction to be set;
2> Press to confirm the setting;
3> Press to cancel the setting.
Impact DIR
< MEAS Setting
Impact DIR
Hard/strength
Tolerance
Hardness Unit
Material
Average Times
- 51 -
The following optional materials are displayed when
【HARDNESS】 is set: steel/cast steel, alloyed tool steel,
stainless steel, gray cast iron, nodular cast iron, cast aluminum
alloys, copper-zinc alloy(brass), copper-tin alloy(bronze),
copper and forged steel.
The following optional materials are displayed when
【STRENGTH】 is set: mild steel, high carbon steel, chrome
steel, chrome vanadium steel, chromium-nickel steel,
chrome-molybdenum steel, Cr-Ni-Mo steel, chromansil,
super-strength steel and stainless steel.
(2) Setting of Average Times
(3) Setting of Material
1> Press to select material upward;
2> Press to select material downward;
3> Press to select material leftward;
4> Press to select material rightward;
5> Press to complete the modification;
6> Press to cancel the modification.
Note: 1. After changing material setting, the hardness unit
automatically converse into HL unit.
2. Set the material first, then set the hardness unit.
User can alter average times in the range of 1~32 times.
1> Press to modify the value;
2> Press to select the number
to be modified;
3> Press to complete the
modification;
4> Press to cancel the modification.
ST&CST CWT ST
SS GC IRON
NC IRON C ALUM
Copper Forged ST
Brass Bronze
Average Times
3
- 52 -
(4) Setting of Hardness Unit
Note:
1. The hardness unit will display only when the selected impact
device and material can be converted, otherwise it will not
display.
2. Set the material first, then set the hardness unit.
3. The hardness unit automatically restores to HL unit after
changing material setting.
Note:
1. Resetting will be reminded if the set value is out of the
measuring range.
2. If the lower limit is larger than the upper limit, they will
automatically swap. The default lower limit is 170HL and the
upper limit is 960HL.
1> Press key to move cursor
over the needed hardness unit;
2> Press key to confirm;
3> Press key to cancel
operation.
(5) Setting of Tolerance Limit
1> Press to modify the value;
2> Press to select the number
to be modified;
3> Press to complete the modification;
4> Press cancel the modification.
Hardness Unit
HL HV HRC
Tolerance
Lower
Upper 0 9 6 0
0 1 7 0
- 53 -
1> Press to select Hardness
pattern or Intensity pattern;
2> Press to complete the modification;
3> Press to cancel the modification.
Enter the menu, select storage
management , and short press
to enter the options, referring to
< 6.4 Menu Operations>.
Operations are as follows:
1> Press to select option upward;
2> Press to select option downward;
3> Press to enter the option;
4> Press to back to the main menu.
(6) Setting of Hardness/Intensity
6.6 Storage Management
Hard/Strength
Hard Strength
MEAS Setting Record
SYS Setting Calibration
Information RES default
08:28
- 54 -
● Browse from the first group
Display storage data from the first group. By default, the
instrument displays the record list from the page where the
first group of data is located;
● Browse from the last group
Display storage data from the last page. By default, the
instrument displays the record list from the page where the
last group of data is located;
For specific data browsing, refer to Page 56 (5) Data Browsing
(1) Browse from the first group/ Browse from the
last group
1> Press to modify the value;
2> Press to select the number
to be modified;
3> Press to display memory data
from the input initial group;
4> Press to cancel the operation.
(2) Browse from the selected group
Select Range
From 001 To 002
< Record
The first item
Delete all
Delete item
The current item
The last item
< Record
The first item
Delete all
Delete item
The current item
The last item
- 55 -
(3) Delete the selected group
(4) Delete all data
1> Press to modify the value;
2> Press to select the number to
be modified;
3> Press to delete the selected group;
4> Press to cancel the operation.
Delete all data: Delete all data in the memory.
1> Press to display the dialog box
【Delete or not?】. Press to
select Yes, and press to delete data.
2> Press to select No. Press
to cancel the deletion operation.
3> In the prompt box page, pressing
can also cancel the deletion operation.
Note:
1. The input data cannot exceed the figure of existing groups.
2. The initial group cannot be larger than the last group.
3. After the selected group is deleted, the serial numbers of all
stored data are rearranged.
4. When deleting data, especially the data of the little group
number, it may take about 30s at most to move the following
data. At this time, please do not shut down the instrument to
avoid data chaos.
Are you sure you
want to delete?
Yes No
Select Range
(001 To 002)
From 001 To 001
- 56 -
● Data menu: The page displays one or more groups of
numbered data
1> Press to select data
groups with different numbers ;
2> Press to enter the data
browsing page of the data group
and browse the data records;
3> Press to exit browsing.
● Data browsing page: The page displays the recorded test data
1> Press to turn the page up and down for more data:
average value, measurement conditions, or single
measurement value;
2> Press to return to the data selection page.
Number:No.001
Model:D
Directio:
Second: 03
Matera:ST&CST
Avg:792HL
Measured value:
788 795 792
(5) Data Browsing
No.001 800HL
No.002 827HL
No.003 878HL
No.004 821HL
No.005 880HL
No.006 876HL
No.007 819HL
No.008 863HL
No.009 850HL
No.010 912HL
No.011 863HL
No.012 850HL
No.013 912HL
- 57 -
6.7 System Setting
System setting menu: Select system setting , press to
enter the options, referring to <6.4 Menu Operations>.
Operations are as follows:
[ Remove error ]: It works by 3σ code to delete values that are
not within the range of (μ-3σ, μ+3σ), μ is the average value,
and σ is the standard difference.
For instance: 16 measurements are made, which are x1, x2,
x3..., x16.
In the formula, Xi means the ith value. If |Xi -μ|>3σ, Xi is
considered to be the largest error, and this value will be deleted.
1> Press to select option upward;
2> Press to select option downward;
3> Auto Save, Remove error, Key Tone,
Alarm Tone and Light Set are directly
set with key, and Brightness,
Language and Date Time Set are set
by pressing the key to enter;
4> Press to return to the previous page.
μ=(x1+x2+...+x16)/16
< SYS Setting
Auto Save:
Remove error:
Key Tone:
Alarm Tone:
Light Set:
ON
OFF
ON
ON
ON
Brightness
Language
Date Time Set
(6) Transfer data
[ Transfer Data ] Send memory data in text format from the
communication port. (Only valid when connecting to a computer)
- 58 -
[ Date Time Set ]:The current time and date are displayed on
the screen in the date format of \"year/month/date\", in the time
format of \"hour/minute/second\".
[ Key Tone ]: When it's ON, the device will \"beep\" once every
time the key is pressed;
[ Alarm Tone ]: When it's ON, the device will \"beep\" once if the
test value is out of the tolerance limit or the data is deleted.
[ Light Set ]: When it's ON, when the measured value exceeds
the tolerance limit, the red light will light up, and when normal,
the green light will light up.
[ Brightness ]:
1. Press to increase brightness;
2. Press to reduce brightness;
3. Press to complete the change;
4. Press to cancel the change.
Brightness setting
Press[�]Brighten
Press[�]Dimming
[ Language ]:Press to select language, press to
save the setting, press to exit.
语言/language
中文 English
- 59 -
1> Press to modify the
value;
2> Press to select the time
and date to be modified;
3. Press to complete the change;
4. Press to cancel the change.
Date Time Set
Date 23 | 12 | 04
Time 15 | 32 | 28
6.8 Software Information
6.9 Instrument Calibration
Enter menu, select software information and short press
to enter options, referring <6.4 Menu Operation>.
Enter menu, select instrument calibration and short press
to enter options, referring <6.4 Menu Operation>.
This interface displays information
about the instrument and embedded
software. Content changes with
software upgrades and is not
updated here.
Version:V1.00202308
Name: HL-Simple
Calibration
0 of 5
Average:
Real:
780
780
0 of 5
Calibration
- 60 -
6.10 Factory Reset
Calibration requires the user to measure 5 times and use the
average value of 5 measurements as the calibration value.
The distance between each measurement is about 3mm. The
average and true values are displayed when the measurement
is completed.
1> Press to modify the value;
2> Press to select the value to be modified;
3> Press to complete the calibration, then short press
to return to the menu;
4> Press cancel calibration.
1> Press to select Yes or No;
2> Press to select Yes, then restore
factory settings, select no to return to
the setting menu;
3> Regardless of the cursor position,
pressing can also cancel the
operation and return to the setting
menu directly.
Enter menu, select factory reset and short press to
enter options, referring <6.4 Menu Operation>.
RES default
Do you want to restore?
Yes No
- 61 -
● The instrument has automatic shutdown function to save
battery power; If there is neither measurement nor any key
operation within 5 minutes, the instrument will shut down
automatically.
● When the battery voltage is too low, the instrument will
display , and then automatically shut down.
● Clean the conduit and the impact body of the impact device
with nylon brush after utilizing 1000~2000 times. First screw
the back-up ring off, take the impact body out, screw nylon
brush inside conduit counterclockwise till the end, then pull it
out. Repeat 5 times, then install the impact body and back-up
ring;
● Release the impact body after using;
● All kinds of lubricants are not allowed inside the impact device.
1. Maintenance of Impact Device
Fault Analysis and Troubleshooting
Maintenance and Repair
Problem Reason Solution
Charge them
in time
Change sensor
cable
Batteries are
exhausted
Sensor cable
malfunction
Fail to turn on
No readings
shown
6.11 Auto-off
- 62 -
● The instrument should be far away from vibration, strong
magnetic field, corrosive media, moisture, dust, and stored at
room temperature;
● Under the condition that the original packaging is maintained,
the transportation can be carried out on the tertiary highway.
● When the standard Rockwell hardness blocks are used for
test and all errors are greater than 2HRC, the possible reason
is that the head is worn out. So the ball head or impact body
should be replaced.
● Please do not disassemble or adjust any fixed parts if there is
any abnormal phenomenon with the device. Please fill in the
warranty card and send it to our maintenance department for
after service.
● The built-in 3.7V 2000mAh lithium battery is non-removable.
Please charge the device in time when it fails to start or the
battery level is low. Please charge it with adapter of DC 5V
1A whose charging port is Type-C.
● When charging, a scrolling battery symbol is displayed. After
a full charge, the battery symbol is displayed as full
power.
Note: When user don't use the device for a long time, charge it
fully and recharge it every six months to avoid battery damage.
2. Repair of Instrument
3. Charging and Maintenance of Li-ion Battery
4. Storage Condition, Transportation and Notes
- 63 -
Material Impact device
D/DC D+15 C G E DL
HRC 17.9~68.5 19.3~67.9 20.0~69.5 22.4~70.7 20.6~68.2
HRB 59.6~99.6 47.7~99.9 37.0~99.9
HRA 59.1~85.8 61.7~88.0
HB 127~651 80~638 80~638 90~646 83~663 81~646
HV 83~976 80~937 80~996 84~1042 80~950
HS 32.2~99.5 33.3~99.3 31.8~102.1 35.8~102.6 30.6~96.8
Forged
steel HB 143~650
Alloyed
Tool steel HRC 20.4~67.1 19.8~68.2 20.7~68.2 22.6~70.2
HV 80~898 80~935 100~941 82~1009
Stainless
steel
HRB 46.5~101.7
HB 85~655
HV 85~802
Gray
Cast iron HB 93~334 92~326
Nodular
Cast iron HB 131~387 127~364
Alloys HB 19~164 23~210 32~168
HRB 23.8~84.6 22.7~85.0 23.8~85.5
(Brass)
CopperZinc Alloys HB 40~173
HRB 13.5~95.3
Copper-tin
Alloys HB 60~290
Copper HB 45~315
Unit
Steel/
Cast steel
Attached Table 1
- 64 -
Attached Table 2
NO. Material HLD
350~522
σ b(MPa)
374~780
737~1670
707~1829
704~1980
763~2007
721~1875
844~1933
755~1993
1180~2652
703~1676
500~710
500~730
500~750
500~750
500~738
540~738
500~750
630~800
500~710
1
2
3
4
5
6
mild steel
high carbon steel
Cr-V steel
Cr-Ni steel
Cr-Mo steel
chrome steel
7 Cr-Ni-Mo steel
8 Cr-Mn-Si steel
9 SSST steel
10 SST steel
- 65 -
Attached Table 3
Shaped impact
device DC(D)/DL D+15 C G E
impact energy 11mJ 11mJ 2.7mJ 90mJ 11mJ
impact body quality 5.5g/7.2g 7.8g 3.0g 20.0g 5.5g
bulb-head hardness 1600HV 1600HV 1600HV 1600HV 5000HV
bulb-head diameter 3mm 3mm 3mm 5mm 3mm
bulb-head material tungsten
carbide
tungsten
carbide
tungsten
carbide
tungsten
carbide adamas
impact device
diameter 20mm 20mm 20mm 30mm 20mm
impact device length 86(147)/
75mm 162mm 141mm 254mm 155mm
impact device quality 50g 80g 75g 250g 80g
sample max
hardness 940HV 940HV 1000HV 650HB 1200HV
bulb-head indentation dimension
sample surface
mean roughness 1.6µm 1.6µm 0.4µm 6.3µm 1.6µm
sample min quality
testing directly
stable support
tight coupling
>5kg
2~5kg
0.05~2kg
>5kg
2~5kg
0.05~2kg
>1.5kg
0.5~1.5kg
0.02~0.5kg
>15kg
5~15kg
0.5~5kg
>5kg
2~5kg
0.05~2kg
sample min thickness
tight coupling
harden layer min depth
5mm
≥0.8mm
5mm
≥0.8mm
1mm
≥0.2mm
10mm
≥1.2mm
5mm
≥0.8mm
hardness 300HV
indentation diameter
indentation depth
0.54mm
24µm
0.54mm
24µm
0.38mm
12µm
1.03mm
53µm
0.54mm
24µm
hardness 600HV
indentation diameter
indentation depth
0.54mm
17µm
0.54mm
17µm
0.32mm
8µm
0.90mm
41µm
0.54mm
17µm
hardness 800HV
indentation diameter
indentation depth
0.35mm
10µm
0.35mm
10µm
0.35mm
7µm
---
---
0.35mm
10µm
the range of impact device application:
DC-type for hole or inner cylindrical surface;
DL-type for spindly narrow groove or hole;
D-type for regular testing;
D+15 for groove or sunken surface due to its smaller contact area and lengthening;
C-type for hardened layer and samples with minor diameter and thinner wall due to
its smaller impact force which usually cause damage on surface;
G-type for larger thickness and more rough casting samples;
E-type for highest hardness materials.
- 66 -
Model No. Remark Shaped back-up
ring sketch No.
1 Z10-15
Z14.5-30
Z25-50
HZ11-13
HZ12.5-17
HZ16.5-30
K10-15
K14.5-30
HK11-13
HK12.5-17
HK16.5-30
UN
test outside cylindrical
surface R10~R15
test outside cylindrical
surface R14.5~R30
test outside cylindrical
surface R25~R50
test inner cylindrical
surface R11~R13
test inner cylindrical
surface R12.5~R17
test inner cylindrical
surface R16.5~R30
test outside sphere
surface SR10~SR15
test outside sphere
surface SR14.5~SR30
test inner sphere
surface SR11~SR13
test inner sphere
surface SR12.5~SR17
test inner sphere
surface SR16.5~SR30
test outside cylindrical
surface, adjustable
radius R10~∞
2
3
4
5
6
7
8
9
10
11
12
Attached Table 4
Note: When measuring cylinder or sphere, please use the
corresponding special-shaped supporting ring to ensure the
measurement accuracy.