- 48 -
In the menu setting state, select \"Factory reset\" and press
to enter the setting page; Press again to confirm
the setting. Press to cancel the setting.
BACK ENTER
MENU
Alarm record
Calibration
Zero offset
Alarm limit
Alarm setting
Date and time
Auto off
Language
Factory reset
BACK ENTER
MENU
Alarm record
Calibration
Zero offset
Alarm limit
Alarm setting
Date and time
Auto off
Language
Factory reset
5.9 Factory Reset
BACK ENTER
Language
中文
ENGLISH
Press
key
BACK ENTER
Factory reset
Are you sure?
Press
key
- 49 -
1. Prevent the instrument from falling from the height or
violent vibration.
2. In the presence of high concentrations of gas, the
instrument may not be used normally.
3. Please strictly follow the instructions to operate,
otherwise the test result may be inaccurate or damaged.
4. The instrument shall not be used in an environment
containing vapor, acidic gas, alkaline gas, strong
oxidizing gas and dust.
5. Do not store or use this product in an environment
containing corrosive substances (such as chlorine,
silicon-containing gas, etc.), and do not use and store
this machine in other harsh environments, including too
high, or too low temperature, high humidity,
electromagnetic field and strong sunlight exposure.
6. If there is dirt on the surface of the instrument, wipe it
gently with a clean soft cloth dipped in water. Do not
use corrosive solvents or hard objects to wipe the surface
of the instrument, otherwise the surface of the
instrument may be scratched or damaged.
7. In order to ensure the detection accuracy, the instrument
should be calibrated regularly, and the calibration cycle
should not exceed 1 year.
8. Please contact us for any application or fault beyond
the description in this manual.
6. Notes
- 50 -
7. Common Problems and Solutions
Problem Reason Solution
Low voltage Charge it in time
Crash Contact dealer or
manufacturer for repair
Circuit fault Contact dealer or
manufacturer for repair
Circuit fault Contact dealer or
manufacturer for repair
Sensor expiration
Contact dealer or
manufacturer to replace
the sensor
Uncalibrated for
long time Calibration in time
Unable to start
No reaction to
the detected gas
Excessive
sensor drift
Calibrate or replace the
sensor in time
Zero setting
failure
Inaccurate
display
Battery dead Charge and reset the
time in time
Strong
electromagnetic
interference
Reset time
Sensor drift Zero calibration
Wrong time
display
Normal detection
non-return to
zero (except O2)
Sensor failure
Contact dealer or
manufacturer to replace
the sensor
The interface
displays full scale
when the
instrument is in
normal test
- 51 -
Please check carefully whether all accessories of the
instrument are complete according to the following list.
If incomplete, please contact the dealer or manufacturer
in time.
The instrument should be stored in a ventilated room with an
ambient temperature of 0 ° C ~ 40 ° C and a relative humidity
of no more than 85%. Avoid direct sunlight, and the air can
not contain harmful gases or impurities that will corrode the
instrument.
8. Storage
9. Packing List
Meter PC 1
No. Name Unit QTY Remark
1
2
3
4
6
PC
User manual PC
1
Hose
1
PC
Color box
Type-C cable PC 1
1
5 Cloth bag PC 1
7 Multi-function
detection handle PC 1
PP box for flagship
Only flagship
Sndway Technology (Guangdong) Co., LTD
全国咨询服务热线/ Service Hotline:400-125-6969
深达威科技(广东)股份有限公司
Add: Sndway Science & Technology Industrial Park, 58
Tuanjie Road, Humen 523930, Dongguan, China
地 址:东莞市虎门镇虎门团结路58号深达威科技园
电 话 / Tel:0769-85265688
网 址/ Web:www.sndway.com
邮 箱/E-mail:market@sndway.com